Introduction: Microinjection is a core tool in the biomedical field, covering both basic research and clinical applications, with use cases including cell biology, systems biology and developmental biology research, stem cell gene manipulation, intracytoplasmic sperm injection (ICSI), disease modeling, and clinical procedures such as in vitro fertilization (IVF), preimplantation genetic diagnosis and intraocular injection. The core carrier of this technique is the hollow microneedle (also called a micropipette or microcapillary needle), whose main function is to deliver trace amounts of substances such as cells, DNA, RNA, proteins, viruses and micro/nanoparticles directly into biological targets such as cells, embryos, tissues and organs. The mainstream fabrication process for conventional hollow microneedles is: heating and softening a glass capillary → axial pulling to form a tapered tip → cooling and solidification → cutting/grinding the tip to form a single opening (inner/outer diameters adjusted as needed). This process determines the inherent structure of conventional microneedles — a single opening only at the very tip, collinear with the insertion direction — which plants the seeds of clogging problems in later applications. Although some studies have used conventional micromachining to fabricate "side-hole microneedles" to reduce clogging, such processes are time-consuming and costly, and have been difficult to popularize in microinjection over the past decades. To address these pain points, researchers at the University of Maryland proposed a dual solution of "structural reconstruction + manufacturing innovation" based on ex situ 3D printing with two-photon direct laser writing (ex situ DLW, esDLW), published under the title "3D nanoprinting of embryo microinjection needles with anti-clogging features" in the authoritative journal Microsystems&Nanoengineering.

Abstract: Biomedical applications spanning research and clinical practice all depend on the microinjection workflow, which uses a slender hollow microneedle to deliver exogenous substances directly into biological targets such as embryos. Unfortunately, conventional microneedles clog easily (for example, during penetration, cytoplasmic material from the embryo becomes lodged inside the tip and obstructs delivery), prompting researchers to modify the tips with top-down micromachining to reduce such failures. Recent advances in submicron additive manufacturing, two-photon direct laser writing (DLW), provide a new bottom-up route to reconstruct tip structures by geometric design and thereby solve the clogging problem. In this study, we fabricated monolithic hollow microneedles by 3D printing (650 µm tall, 15 µm in diameter) that integrate three structural features designed to mitigate clogging: (1) a solid, fine tip; (2) multiple side holes (i.e., perpendicular to the insertion direction); and (3) a built-in microfilter. Serial microinjection experiments in live zebrafish embryos showed that these 3D-printed microneedles deliver more effectively and exhibited no complete clogging events, whereas complete clogging was very common in both standard glass microneedles and 3D-printed control microneedles. These results demonstrate the unique potential of DLW-based 3D printing for high-precision microinjection applications, especially in scenarios requiring large numbers of serial injections or involving critical delivery substances and targets.



Innovations:

Based on ex situ 3D printing with two-photon direct laser writing (DLW) (esDLW), the study proposes solutions along two dimensions — "structural reconstruction" and "manufacturing innovation". The core innovations are as follows:


Anti-clogging structural design (triple cooperative protection): for the first time, three anti-clogging features are integrated into a monolithically fabricated hollow microneedle, addressing clogging from three levels: prevention, interception and redundancy.

Solid fine tip: it replaces the conventional tip opening, preventing target substances from entering the needle directly during penetration and reducing the causes of clogging at the source;

Multiple side-hole array: 20 side holes 5 µm in diameter (perpendicular to the insertion direction) are arranged below the tip, so that delivery fails only when all side holes clog simultaneously, greatly reducing the probability of complete clogging;

Built-in microfilter: a 3.25 µm×3.25 µm pore array is integrated at the base of the microneedle to intercept particles/aggregates in the fluid in advance and prevent clogging of the downstream channel.

Manufacturing breakthroughs:

Monolithic fabrication and seamless sealing: the microneedle structure is printed directly on the end of a glass capillary by esDLW without any adhesive, achieving a fully fluid-tight seal between the microneedle and the capillary and preventing fluid leakage during injection;

High precision and high repeatability: printed microneedles have outer diameter (OD) errors of only ±0.1~0.2 µm and inner diameter (ID) errors of ±0.1~0.2 µm, far better than conventional glass needles (OD error ±4.0 µm, ID error ±3.1 µm), and can be used directly without post-calibration;

High efficiency and low cost: printing an anti-clogging 3D microneedle takes only 10 minutes, versus 5 minutes for the control needle, avoiding the high cost and long cycle of conventional micromachining and making it suitable for large-scale applications.

玉之泉

Figure 1 esDLW fabrication workflow, microneedle structure and zebrafish injection principle

a-c: esDLW fabrication workflow

a: a glass capillary (amber fused silica, ID=75 µm, OD=360 µm) is loaded into the DLW printer, with its tip coated with IP-L photoresist;

b: a femtosecond infrared laser scans "point by point and layer by layer" through the photoresist, forming the microneedle structure by two-photon polymerization (2PP);

c: after printing, the microneedle and the capillary form a fully fluid-tight integrated body without adhesive.

d: Core structure of the 3D anti-clogging microneedle

The three anti-clogging features — "solid fine tip", "multiple side-hole array" and "built-in microfilter" — are labeled, visually showing the structural difference from conventional microneedles.

e: Fluid delivery path

Shows the complete path of the fluid from the capillary → microfilter (particle interception) → main channel → side holes (outflow), explaining the "filter-side hole" cooperative anti-clogging mechanism.

f-i: Zebrafish embryo injection steps

f: the solid fine tip penetrates the embryo surface;

g: the microneedle penetrates the embryo (with no risk of tip-opening clogging);

h: fluid is delivered into the embryo through the side holes;

i: the microneedle is withdrawn, leaving the delivered substances in the embryo to complete the injection.

玉之泉

Figure 2 Fabrication results of esDLW-printed microneedles

a-c: 3D anti-clogging microneedle

a: computer-aided manufacturing (CAM) simulation showing the overall microneedle structure (solid tip + side holes + filter);

b: bright-field microscope image during printing, showing the microneedle forming in real time;

c: SEM image (with local magnification, scale bars 100 μm/10 μm), clearly showing the seamless connection between the 20 side holes of 5 μm and the capillary interface, with no physical defects.

d-f: esDLW control microneedle

d: CAM simulation (single opening at the tip, no side holes);

e: bright-field image during printing;

f: SEM image (scale bars 100 μm/10 μm), verifying the tip opening structure and dimensional accuracy of the control needle (OD≈15.4 μm).

g-h: Characterization of the built-in microfilter

g: cross-sectional SEM image of the microneedle (scale bar 100 μm, with 25 μm local magnification);

h: SEM image of the underside of the microfilter (scale bar 25 μm), showing uniform 3.25 μm×3.25 μm micropores and demonstrating its ability to intercept particles.

玉之泉

Figure 3 Results of microfluidic burst pressure experiments

a-b: Fluid injection verification

a: bright-field image of a control needle injecting rhodamine-stained DI water;

b: fluid injection with the 3D anti-clogging needle; both show that fluid flows out only from the designed openings (tip/side holes) with no leakage at the interface (scale bar 100 μm).

c-d: Linear pressure-flow curves

c: control needle (n=10);

d: 3D anti-clogging needle (n=10); the curve is stably linear (error bars indicate standard deviation) with no abnormal fluctuations, proving that the interface neither leaks nor ruptures from 0 to 200 kPa.

e-f: Cyclic pressure-flow curves

e: control needle;

f: 3D anti-clogging needle (100 cycles of 110 kPa/0 kPa); the flow rate recovers to a stable level after each pressure load (no "flow drift"), proving that the interface withstands the repeated loads of serial injection.

玉之泉

Figure 4 Fluorescence intensity results of in vivo injection into zebrafish embryos

a-d: Micrographs of the injection process

a, c: bright-field images during injection with the 3D anti-clogging needle (a) and after injection (c);

b, d: the corresponding fluorescence images, showing rhodamine-stained fluid successfully delivered into the embryo (scale bar 250 μm).

e: Bar chart of fluorescence intensity statistics

The horizontal axis shows the injection groups (1~20, 21~40 ... 81~100 injections) and the vertical axis the fluorescence intensity (a.u.); error bars indicate standard deviation (n=10 needles per type, 3,000 injections in total). The results show that the fluorescence intensity of the 3D anti-clogging needle is significantly higher than that of the conventional needle (p<0.0001) and the control needle (p<0.001), demonstrating higher delivery efficiency.

玉之泉

Figure 5 Clogging results of zebrafish embryo injection

a: Schematic of the injection process

Shows serial injection of zebrafish embryos with the 3D anti-clogging needle (scale bar 500 μm).

b-d: Normalized delivered volume curves

The horizontal axis shows the injection number (0~100) and the vertical axis the normalized delivered volume (compared with the mean of the initial dummy injections):

b: conventional glass needle, with a significant decrease in volume (lowest 0.30±0.37, p<0.0001) and large fluctuations (standard deviation 0.37~0.51);

c: esDLW control needle, significantly lower at only 2 points (60/80 injections, p<0.05/0.01), with smaller fluctuations than the conventional needle;

d: 3D anti-clogging needle, slightly lower only at the 100th injection (0.85±0.20, p<0.05), with the smallest fluctuations (standard deviation 0.07~0.26).

e-g: Statistics of complete clogging events

The horizontal axis shows the microneedle number (1~10) and the vertical axis the number of complete clogging events:

e: conventional needles; 7 needles clogged 3 or more times, with an overall clogging rate of 44.0±26.3%;

f: control needles; 2 needles clogged 3 or more times, with an overall clogging rate of 26.0±23.2%;

g: 3D anti-clogging needle; no complete clogging events occurred at all, indicating a significant anti-clogging effect.

Addressing the pain points of conventional embryo microinjection needles — easy clogging, large dimensional variation and hard-to-adopt alternatives — the study developed a hollow microneedle with triple anti-clogging features based on esDLW 3D printing. Its core contributions are as follows:

Technological breakthrough: for the first time, the "solid fine tip + multiple side holes + built-in microfilter" design is fabricated monolithically; combined with the high precision (dimensional error <0.2 μm) and high efficiency (printing <10 minutes) of esDLW, it resolves the inherent contradictions of conventional manufacturing;

Performance validation: microfluidic tests prove that the interface withstands 200 kPa of linear pressure and 100 pressure cycles without leakage; zebrafish embryo experiments show no complete clogging events with the 3D anti-clogging needle, with delivery efficiency (fluorescence intensity) more than doubled compared with conventional needles and possibly reduced embryo damage;

Application value: the technique applies not only to embryo microinjection but can also be extended to delivery into other biological targets such as cells and tissues, as well as to drug delivery, lab-on-a-chip systems and other fields, providing a reliable microneedle solution for biomedical research and clinical procedures.

A limitation of the study is that post-operative embryo survival was not directly evaluated; future work can further optimize the biocompatibility of the microneedle material and extend injection validation to larger biological targets (such as mammalian embryos).



Source:Sarker, S., Wen, Z., Acevedo, R. et al. 3D nanoprinting of embryo microinjection needles with anti-clogging features. Microsyst Nanoeng 11, 171 (2025).

Article URL:https://www.nature.com/articles/s41378-025-01005-2